SECcurity Detectors 



SECcurity column compartment 80°/100°
  SECcurity GPC1200 Refractive Index RI detector  
Part number: 404-0006 
The detector of  choice for the anaylsis of non-UV absorbing samples. Ideal concentration detector for the combination with on-line light scatterging detectors or on-line viscometers. High signal-to-noise ratio is enabled by a minimum baseline noise. Electronic temperature regulation maintains a steady optical unit temperature up to 55° C. Convenient early maintenance feedback (EMF) is available. An automatic solvent recycle valve is already included. 

Specifications    
Detection type:  Deflection mode 
Short-term noisehort-term noise:  2.5 x 10-9 RIU
Drift:  200 x 10-9 RIU/h
Refractive index range:  1.00–1.75, calibrated
Flow cell:  8 µL, 5 bar pressure maximum
Temperature control  Ambient +5 °C to 55 °C
pH range:    2.3 - 9.5
Time programmable   Polarity, peak width
Zero adjustment   Automatic zero
Valves:   Automatic purge and automatic solvent recycle
Analog output:  Recorrder/integrator: 100 mV or 1 V, with offset   adjustment, RIU range selectable
GLP/Compliance:  Early maintenance feedback (EMF) for continuous tracking of instrument usage with user-settable limits (purge interval) and feedback messages. Electronic records of maintenance and errors, automated OQ/PV procedures.



SECcurity ELS Detector
  SECcurity Evaporative light scattering detector ELS4000
Part number: 409-0025
ELSD with superior and sensitive detection for a wide variety of chemical classes and structures. With ThermoSplit technology for the optimal signal to noise ratio.
Low detection limits down to 10 ng. Low evaporation temperatures and outstanding reproducibility. No baseline shift even with extreme gradients.  The fully warranted Teflon nebulizer provides superior solvent and chemical resistance, eliminates sample build-up even at high temperatures, does not permit sample carry over or increased noise from surface leaching. Reproducible droplet distributions are assured by thermally isolating the nebulizer from the spray chamber and drift tube, improving reproducibility between methods.
Low gas pressure and flow requirements allows use of cylinder, house systems, or gas generators. The pneumatic system is protected by a high pressure shut off, and provides a signal in the event of unstable or low pressure.

Specifications
Detector type:  Evaporative light scattering detector with Thermo-Split 
Evaporative Zone Temperature:  25° C - 120° C
Thermo-Split™ Chamber Temperature:  10° C - 70° C
Light source 670 nm laser diode, <5 mW
Flow rate:  0.2ml/min to 5ml/min
Detection limit:  10 ng
Gas requirements:  4 - 5,5 bar (60-80psi) nitrogen or other inert gas
Analog output:  0 - 5 VDC



Variable Wavelength detector
  SECcurity Variable wavelength UV-VIS
Part number: 409-0007
The SECcurity GPC1200 variable wavelength detector (VWD) lowers detection limits with simplified optics for high sensitivity, low baseline drift and wide linear range. Achieve high sensitivity and low detection limits with low baseline noise and drift. Cut operating costs with automatic shutdown for longer productive lamp lifetime.
Specifications
Detector type:  Double-beam photometer
Light source Deuterium lamp

Drift  3 x 10-4AU/hr, at 254 nm
Linearity  >2 AU upper limit
Wavelength range:  190-600 nm
Wavelength accuracy: ± 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter 
Wavelength bandwidth: 6.5 nm typical over whole wavelength range
Analog output: Recorder/integrator 100 mV or 1 V
Flow cells:  
Standard: 14-µL volume, 10-mm cell path length and 40 bar (588 psi) pressure maximum
High Pressure: 14-µL volume, 10-mm cell path length and 400 bar (5880 psi) Pressure maximum
Micro: 1-µL volume, 5-mm cell path length and 40 bar (588 psi) pressure maximum
Semimicro: 5-µL volume, 6-mm cell path length and 40 bar (588 psi) pressure maximum
GLP/Compliance: Electronic records of maintenance and errors. Early maintenance feedback (EMF) for continuous tracking of instrument usage, display of feedback messages if preset limits are exceeded. Verification of wavelength accuracy with built-in holmium oxide filter.

SECcurity GPC1200 Multiple Wavelength Detecto
  SECcurity Multiple wavelength UV-VIS
Part number: 404-0008
The SECcurity GPC1200 MWD is a dedicated multiple wavelength detector that offers the same high sensitivity for both single and multiple wavelength detection. 
Diode array design enables simultaneous acquisition of up to 8 compound-specific wavelengths for increased sensitivity and selectivity. Deuterium and tungsten lamps with a wavelength range of 190 to 950 nm offer an expanded detection range with lowest detection limit.
Specifications
    
Detector type:  1024-element diode array
Light source:  Deuterium and tungsten lamp
Number of signals:  8
Short-term noise:  ± 0.8 x 10 -5 AU at 254 nm and at 750 nm
Drift:  0.9 x 10 -3 mAU/h at 254 nm
Linearity: >2 AU upper limit
Wavelength range:  190-950 nm
Wavelength accuracy:  ± 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter
Flow cells:
standard: 13-µL volume, 10 mm-cell path length, 120 bar (1760 psi) pressure maximum
micro: 2-µL, volume, 3-mm cell path length, 400 bar (5880 psi) pressure maximum
preparative: 3-mm cell path length, 120 bar (1760 psi) pressure maximum
0.3-mm cell path lenght, 20 bar (294 psi) pressure maximum, quartz
0.06-mm cell path length, 20 bar (294 psi) pressure maximum, quartz
others on request
Analog output: Recorder/integrator 100 mV or 1 V, 2 outputs
GLP/Compliance: Electronic records of maintenance and errors. Early maintenance feedback (EMF) for continuous tracking of instrument usage, display of feedback messages if preset limits are exceeded. Verification of wavelength accuracy with built-in holmium oxide filter.


SECcurity GPC1200 Diode Array Detector
  SECcurity Diode Array Detector (DAD)
Part number 404-0010
WinGPC Unity SystemPilot  Module
The SECcurity GPC1200 DAD with improved noise specifications enables scientist to reach lower detection limits, even under harsh and fluctuating ambient temperature and humidity conditions.
Deuterium and tungsten lamps with a wavelength range of 190 to 950 nm offer an expanded detection range with lowest detection limit. 1024 diodes and a 1-nm slit achieve high spectral resolution.
  
Specifications
Detector type:  1024-element diode array
Light source:  Deuterium and tungsten lamp
Number of signals:  8
Short-term noise: ± 0.8 x 10 -5 AU at 254 nm and at 750 nm
Drift:  0.9 x 10 -3 AU/h at 254 nm
Linearity: >2 AU upper limit
Wavelength range:  190-950 nm
Wavelength accuracy  ± 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter
Flow cells
Standard: 13-µL volume, 10 mm-cell path length, 120 bar (1760 psi) pressure maximum
Micro: 2-µL, volume, 3-mm cell path length, 400 bar (5880 psi) pressure maximum
Preparative: 3-mm cell path length, 120 bar (1760 psi) pressur maximum
0.3-mm cell path lenght, 20 bar (294 psi) pressure maximum, quartz
0.06-mm cell path length, 20 bar (294 psi) pressure maximum, quartz
others on request
Analog output:  Recorder/integrator 100 mV or 1 V, 2 outputs
GLP/Compliance:  Electronic records of maintenance and errors. Early maintenance feedback (EMF) for continuous tracking of instrument usage, display of feedback messages if preset limits are exceeded. Verification of wavelength accuracy with built-in holmium oxide filter

Latest News

  • Liquid Adsorption Chromatography Seminar April 2017 +

    PSS USA Inc. will be presenting a seminar on Liquid Adsorption Chromatography in Amherst, MA on April 11th, 2017.   This course Read More
  • Speaking at 16th Annual Csaba Horvath Memorial Award Symposium +

    Dr. Derek Lohmnn will be speaking at the 16th Annual Csaba Horvath Memorial Award Symposium on the Yale West Campus in Read More
  • Pullulan ReadyCal Kits for GPC/SEC calibration +

    The calibration of a GPC/SEC-system is a time-consuming process as it requires the accurate weighing of a larger number of different Read More
  • New Column Selector App +

    Compatible with PC, Tablet and Smart Phones The PSS Column Selector helps you to find the right GPC/SEC/GFC columns and Read More
  • WinGPC UniChrom Update v.8.2 +

    WinGPC UniChrom Software Update Available PSS has released the latest version of its macromolecular chromatography data system WinGPC GPC/SEC software. WinGPC UniChrom 8.2 is Read More
  • 1

catalogleft