Search Columns, Standards and Services:

SECcurity manual injectors
The SECcurity GPC1200 offers injection systems for a variety of tasks, ranging from general GPC/SEC needs up to dedicated solutions for higher sample throughput or semi-preparative separations. All modules are designed for quick startup, automated validation and ease of use.

Click picture for close up view

SECcurity manual injector
 
SECcurity manual injector 
Part number: 404-0014
Convenient modular accessory, dimensioned to match the footprint of the SECcurity GPC1200. Stack to create a compact, affordable manual injection GPC/SEC system. Equipped with 100µL samle loop and position sensor. Sample loops from 20µL to 1mL are also available.



SECcurity stadard autosampler
 
SECcurity standard autosampler
Part number: 404-0004
The SECcurity GPC1200 standard autosampler is designed for the reliability, safety, and ease of use needed in (pharmaceutical) routine tasks and quality control, as well as for R&D tasks.
Injection with highest precession allows reliable detection and evalulation with molar mass sensitive detectors and concentration detectors alike.

Specifications   
Injection range: 0.1- 100 µL in 0.1 µL increments
Up to 1500 µL with multiple draw (hardware modification required)

Precision   <0.25% RSD from 5-100 µL, <1% from 1-5 µL, variable volume
Sample viscosity range: 0.2 - 5 cp
Sample capacity: 100 × 2-ml vials in 1 tray; 40 × 2-ml vials in ½ tray
Injection cycle time: Typically 50 s depending on draw speed and injection volume
Carry-over:  Typically < 0.1 %, < 0.05 % with external needle cleaning

Option 900µL enhanced injection kit P/N 409-9015
Specs with 900µL metering device installed
Injection range:  0.1 - 900 µL in 0.1 µL increments (recommended 1 µL increments)
Up to 1800 µL with multiple draw (hardware modification required)
Precision:  Typically < 0.5 % RSD of peak areas from 5 - 2000 µL
Typically < 1 % RSD of peak areas from 2000 - 5000 µL
Typically < 3 % RSD of peak areas from 1 - 5 µL
Variable volume
Sample viscosity range:  0.2 - 5 cp
Sample capacity:  100 × 2-ml vials in 1 tray;  40 × 2-ml vials in ½ tray
Injection cycle time: Typically 50 s depending on draw speed and injection volume
Carry-over:  Typically < 0.1 %, < 0.05 % with external needle cleaning
Pressure operating range:  0-20 MPa (0-200 bar, 0-2950 psi
GLP/Compliance:  Early maintenance feedback (EMF) for continuous tracking of instrument usage with user-settable limits and feedback messages. Electronic records of maintenance and errors, automated OQ/PV procedures.



The standard autosampler can be upgraded from non-thermostatted version to thermostatted version by adding the thermostat
 
Part number: 404-0042

back