Molecular Weight Sensitive Detectors:
Light scattering detector
Capillary viscometer

Click picture for close up view


SECcuritySLD7000
 
SECcurity on-line multi angle light scattering detector
Part number: 403-0001
Software for batch MALLS: Part number: 403-0004
The SECcurity MwA- MALLS is a 7-angle static light scattering detector for the measurement of absolute molecular weight, size, and conformation of macromolecules in solution. It is manufactured by Brookhaven Instruments

The revolutionary design with a cylindrical, self-cleaning, and index-matching cell avoids common pitfalls in GPC/SEC-light scattering detection. Modern ultra-sensitive CCD detection results in high sensitivity, even at low concentration and molar masses.

A small footprint allows integration into any GPC/SEC or FFF system.

The Molecular Weight Analyzer can be used in batch mode (off-line) without modifications.


Specifications 
  
Number of angles:  7 plus additional reference, simultaneous detection
Detection angles/scattering angles:  35, 50, 75, 90 105, 130, 145;
no difference detection angle/scattering angle due to index-matching cell
Cell:  PEEK
Cell volume:  100 µL nominal
Scattering volume:  20 nL nominal
Cell cleaning option:  self cleaning cell with vertical flow path
Light source:    35 mW, 635 nm, vertically polarized, temperature stabilized with precision power control
Detector   fibre optics with low numerical aperture, integral to cell; CCD with ultra-high sensitivity and spatial uniformity
Control    Integrated, dedicated, powerful microcontroller
Input:  4 analog ins with 16 and 24 bit resolution for any concentration detector or viscometer; autosampler/manual injector digital inject in
Options:  temperature control P/N 403-0005
other cell material on request
IR laser or different laser wavelength on request
15 additional analog channels on request
Software for static light scattering measurements in batch mode P/N 403-0004
GLP/Compliance:  fan on/off with built-in temperature control




ETA 2010 online viscometer
 
SECcurity on-line viscometer ETA2010
Part number: 402-0001
The SECcurity eta2010 viscometer four capillary viscometer lowers the detection limit with an asymmetric bridge design. While other viscometers dilute 50% of the sample in the solvent path, the eta2010 dilutes only 20% of the sample in the solvent path.

 80% of the sample run through the sample path leading to an unparalleled signal/noise ratio without any digital signal processing. Low-noise electronics combined with high-sensitivity pressure transducers results in excellent peak shapes and high signal linearity. The eta2010 can be used with any GPC/SEC system. The use of reservoirs in the bridge eliminate viscometer modifations when the column setup changes and reduces the analysis time to 50% compared with other viscometers.

The SECcurity eta2010 can be integrated into any GPC/SEC or FFF system.


Specifications
Detection type:  4 capillary on-line viscometer
Signals: simultaneous measurement delta pressure and inlet pressure ll
Noise level:   0,2 Pa, diff. pressure channel, 25° CK
Cell:  asysmmetric bridge with 80/20 split
Pressure transducers:  fast-responding Hastelloy pressure transducers without hysteresis
Temperature range:  ambient to 80° C, high temperature model to 150° C
Temperature stability:   > 0.01° C
Temperature accuracy:  ± 0.5° C
Maximum flow ratel    1.2 mL
Analog in:  2 analog ins for any concentration detector; Start, Error, Flush, Zero
Analog output:  10.0 Volt/10 kPa FSD Differential Pressure
10.0 Volt/100 kPa FSD Inlet Pressure
GLP/Compliance:    Internal error logbook, electronic pressure transducer protection with error code